10.0 Fuses and Power Settings
10.0
Fuses and Power Settings
10.1
Fuses
F1 - F3: Refer to the circuit diagrams.
F1
Internal
Supply
Fuses
Fitted if supply cable fitted.
Fitted on board to some types
of EMC filter.
GEC Safeclip of the type shown
(glass type F up to 16 A)
38 mm x 10 mm type F fitted on
EMC filter circuit board(s)
F2
Auxiliary
Circuit Fuses
Fitted on board to some types
of EMC filter.
May be omitted up to 25 Amp/
phase supply rating.
2 Amps glass type F
On board: 20 mm x 5 mm
Other: 32 mm x 6 mm
F3
Heat Light
Fuses
May be omitted up to 25 Amp/
phase supply rating.
2 Amps glass type F
32 mm x 6 mm
Customer
Fuses
Required if no supply cable
fitted.
Recommended if cable fitted.
See rating label for current;
See table below for fuse rating.
Model
Phases
Volts
Supply Fuse Rating
MTF 12/25/250
1-phase
110-120 V 10 A
MTF 12/25/250
1-phase
220-240 V 4 A
MTF 12/25/400
1-phase
110-120
10 A
MTF 12/25/400
1-phase
220-240 V 4 A
10.2
Power Settings
Model Volts:
200 V
208 V
220 V
380 V
230 V
400 V
240 V
415 V
all 208 V models
100
100
89
82
75
all other models
100
100
100
100
100
Please refer to the rating label for product specific information.