2.0 Installation
If a metal work tube is being used in the furnace, ensure that it is earthed. In the case
of woundtube models it should be mounted so as not to touch the furnace ceramic tube.
See the 5.3.
Key
A
Tube
B
Insulating Plug
C
Stem
Fig 1 - Insulating Plug
(standard length tube)
2.4
Electrical Connections
Connection by a qualified electrician is recommended.
This product requires a single-phase A.C. supply with earth (ground), which may be Live
to Neutral non-reversible (polarised), Live to Neutral reversible (non-polarised), or Live
to Live.
Check the product rating label before connection. The supply voltage should agree with
the voltage on the label and the supply capacity should be sufficient for the current on
the label.
The supply should be fused at the next size equal to, or higher than the current on the
label. A table of the most common fuse ratings is also given towards the back of this
manual. When the mains cable is factory fitted, internal fuses are also fitted. It is
essential that the operator ensures that the product is correctly fused.
Products with a factory fitted supply cable are designed to be wired directly to an
isolator or fitted with a line plug.
8