2.0 Installation
Depending on the application of the product, it may be appropriate to position it under
an extraction hood. Ensure the extraction hood is switched on during use.
Ensure that the product is placed in such a way that it can be quickly switched off or
disconnected from the electrical supply.
Under no circumstances should any objects be placed on top of the product.
Always ensure that any vents on the top of the product are clear of any
obstruction. Always ensure all cooling vents and cooling fans (if fitted) are
clear of any obstruction.
If the product is supplied with a work tube then the holes at the end of the product
chamber should be supplied at the correct size. Carefully line up the tube and ease it
through the chamber.
For a tube of customer supply, ensure that there is approximately 1 mm free play at
each end; if the tube is too tight, it may crack when heated. Increase the diameter of
the hole by rotating the tube, which is abrasive. Remove loose powdered ceramic fibre
with a vacuum cleaner and see section 7.0.
Contact Carbolite Gero if the holes in the product chamber are too large.
Ensure that the tube is centrally placed and reaches to or through the product end
covers, so that there is no access to exposed electrical connections.