7.6
Element Fitting and Replacement
Replacements: see section 6.7. If at any time the power limit has been increased, reset
it to its original value (see section 10.0 at the back of this manual for the original value).
New elements must not be mixed with aged elements. If a single element fails in an
aged set of elements then replace with a new set. The remaining aged elements can be
used as spare parts in the future.
Disconnect the product from the electrical supply.
Remove the products end guards to gain access to the element connections.
Remember that the elements are fragile and expensive and that they can be damaged
by contamination: handle them with care and keep them clean.
Replacements: make a note of the actual braid and cable connections to the elements.
For rod elements disconnect the clips, lift off the braids and carefully withdraw each
element. Refer to the image below.
Carefully insert the new elements into the product. Ensure that the rod elements are
placed centrally: the same length should be outside the insulation at each end.
Replace the products end guards and connect the product to the electrical supply.
If aged elements have been replaced, reset the products power limit.
If you have any problems with this procedure, please contact the Carbolite Gero
Service.
Element Fitting - Figures
Use of Clip Tool
Element Connections - 6 elements, 2 circuits