2.0 Installation
Supply
Terminal Label
Cable Colour
Supply Types
Live - Neutral
Reversible or Live-
Live
1-phase L
Brown
to live
to either power
conductor
(For USA 200-240V,
connect L1)
N
Blue
to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE
Green/ Yellow
to earth
(ground)
to earth (ground)
Electrical Connection Details:
2- or
3-phase
L1
Black
to phase 1
L2
Black
to phase 2
L3
Black
to phase 3 (
except 2-phase
)
N
Light Blue
to neutral (
except delta
)
PE
Green/ Yellow
to earth (
ground
)
Technical Notes
This model has an electronically controlled power limit. Do not attempt to calculate the
Amps from the Wattage. High break capacity fuses should be used. Avoid fast-blow
fuses and magnetic trip circuit breakers - consult Carbolite Gero if in doubt.