8.0 Fault Analysis
B.
Product Overheats
1.
Product only heats up
when the instrument
switch is ON
The controller
shows a very high
temperature
The controller is faulty
The controller
shows a low
temperature
The thermocouple may be
faulty or may have been
removed out of the heating
chamber
The thermocouple may be
connected the wrong way
around
The controller may be faulty
2.
Product heats up
when the instrument
switch is OFF
The SSR has failed
"ON"
Check for an accidental wiring
fault that could have
overloaded the SSR
38
Summary of Contents for CARBOLITE GERO STF 16/180
Page 4: ...User Power Setting Adjustments 42 11 0 Specifications 43 11 1 Environment 43 4 ...
Page 6: ...1 0 Symbols and Warnings Caution Double Pole Neutral Fusing 6 ...
Page 22: ...3 0 2416 Controller 3 6 Navigation Diagram 22 ...
Page 44: ......
Page 45: ...Notes Service Record Engineer Name Date Record of Work ...