temperature. The rate of cooling is quite fast because of the air inlet blower which is
incorporated in the oven. When the chamber temperature reaches 80 °C the second
dwell period starts and the product remains at 80 °C for 12 hours. At the end of 12
hours the heaters are switched off and the product cools naturally to room
temperature.
5.3
Over-Temperature Control (if fitted)
The over-temperature controller should typically be set at 15 °C above the main
controller. If an over-temperature condition occurs, check the main controller is
functioning correctly.
An over-temperature condition cuts off power to the heating elements. A light in the
over-temperature controller flashes. To reset this, refer to the over-temperature
control section of this manual.
5.4
Solenoid Valve with Manual Switch (if fitted)
If ordered the solenoid valve is operated using the panel mounted
switch. When the switch is in the 'ON' position the solenoid valve will
allow gas to
flow. Ensure the installation and use of the product does not create a hazardous
atmosphere. The workspace must have sufficient ventilation.
13
5.0 Operation