Take care not to disconnect the wires leading from the EMC filter without first recording
their positions: they must be reconnected to the correct terminals.
7.8
Solid-State Relay Replacement
Disconnect the product from the power supply and remove the
appropriate cover as given above.
Make a note of the wire connections to the solid state relay and disconnect them.
Remove the solid state relay from the base panel or aluminium plate.
Replace and reconnect the solid state relay ensuring that the bottom of it has good
thermal contact with the base panel or aluminium plate.
Replace the access panel.
19
7.0 Repairs and Replacements