4.0 3216 Controller
4.7
Programming
4.7.1
Creating a Program
Programs can be created in level 1 or level 2 of the 3216P1 and 3216P5. Each program
contains 8 Ramp/ Dwell pairs.
Note: A currently active program cannot be altered. Go into 'Reset' mode before
starting to create or modify a program
4.7.2
Program Number (3216P5 Only)
Press scroll
until the display shows PROG <PROGRAM NUMBER.> Select the program
number.
4.7.3
Ramp Units
Press scroll
until the display shows RAMP.U <Ramp Units>: Use the up
down
to
select the Ramp Units of Hour, Min or Seconds.
4.7.4
Dwell Units
Press scroll
until the display shows DWEL.U <Dwell Units>: Use the up
down
to
select the Dwell Units of Hour or Min.
4.7.5
Holdback
see section 4.6. Press scroll
until the display shows H.BACK <Program Holdback>: If
a holdback value is required, use the up
down
to enter the required value or
select “OFF” if no holdback is required.
4.7.6
Ramp Rate
Press scroll
until the display shows RMP.1 <Ramp Rate 1>: Using the up
down
enter the value for the first ramp rate depending on the Ramp Units selected. If the
Ramp segment is not required select “OFF”.
4.7.7
Target Setpoint
Press scroll
until the display shows T.SP 1 <Target SP 1>: Enter the temperature
that you want the product to ramp up to using “Ramp rate 1”.
If “Ramp Rate 1” has been set to “OFF”. The product will Heat/ Cool directly to the
Target SP.
4.7.8
Dwell Time
Press scroll
until the display shows DWELL. 1 <DWELL TIME 1>: Enter the time to
dwell at “Target SP 1”. If the dwell segment is not required, use the down
to select
OFF, which is below the Zero value.
This Process is repeated for each of the 8 segments of the program.
20