2.0 Installation
Supply
Terminal Label
Cable Colour
Supply Types
Live - Neutral
Reversible or Live-
Live
1-phase L
Brown
to live
to either power
conductor
(For USA 200-240V,
connect L1)
N
Blue
to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE
Green/ Yellow
to earth
(ground)
to earth (ground)
Electrical Connection Details:
2- or
3-phase
L1
Black
to phase 1
L2
Black
to phase 2
L3
Black
to phase 3 (
except 2-phase
)
N
Light Blue
to neutral (
except delta
)
PE
Green/ Yellow
to earth (
ground
)
DO NOT connect a model ordered for three phase use to a single phase supply or to
the wrong type of three phase supply.
Technical Notes
This model has an electronically controlled power limit. Do not attempt to calculate the
Amps from the Wattage. High break capacity fuses should be used. Avoid fast-blow
fuses and magnetic trip circuit breakers - consult Carbolite Gero if in doubt.
12