5.0 Operation
5.0
Operation
5.1
Operating Cycle
This product is fitted with an instrument switch which cuts off power to the control
circuit.
Connect the product to the electrical supply.
Turn on the instrument switch to activate the temperature controllers. The controllers
illuminate and go through a short test cycle.
Over-Temperature option only
. If the digital over-temperature option has not yet
been set as required, set and activate it according to the over-temperature controller
instructions.
Switch on the heater switch, located on the instrument panel. Unless a time switch is
fitted and is switched off, the product will start to heat up. The Heat light(s) glow
steadily at first and then flash as the product approaches the desired temperature or a
program setpoint.
Over-Temperature option only.
If the over-temperature circuit has tripped, an
indicator on the over-temperature controller flashes and the heating elements are
isolated. Find and correct the cause before resetting the over-temperature controller
according to the instructions supplied.
To switch off power to the heating elements, use the heater switch. To switch the
product off, use both the heater switch and the instrument switch. If the product is to be
left switched off and unattended, isolate it from the electrical supply.
5.2
General Operating Advice
Heating element life is shortened by overheating. Do not leave the
product at high temperature when it is not required. The maximum
temperature is shown on the product rating label and in section 11.0
towards the back of this manual.
Lightweight ceramic fibre insulation can easily be marked by accidental
contact. Some fine cracks may develop in the surface of the insulation
due to the progressive shrinkage of the insulation materials. Cracks are
not usually detrimental to the functioning or the safety of the product.
Clean up any spillages in the insulation, as these can increase the rate of degradation of
the insulation material.