A.

 

Furnace Does Not Heat Up 

 

 

 

 

 

 

1. 

The 

HEAT

 light 

 is 

ON

 

  The heating 

element has failed 

  Check also that the SSR is working 

correctly 
 

2. 

The 

HEAT

 light  

is 

OFF

 

  The controller 

shows 

a very high 

temperature

 or a 

code such as S.br 
 

  The thermocouple has broken or has a 

wiring fault 

 

 

  The controller 

shows a

 low 

temperature

 

  The door switch(es) (if fitted) may be 

faulty or need adjustment 

 

 

 

 

  The contactor (if fitted) may be faulty 

 

 

 

 

 

  The SSR could be failing to switch on due 

to internal failure, faulty logic wiring from 
the controller, or faulty controller 
 

 

 

  There are no lights 

glowing on the 
controller 

  Check the supply fuses and any fuses in the 

furnace control compartment 

 

 

 

 

  The controller may be faulty or not 

receiving a supply due to a faulty switch or 
a wiring fault 

 

 

 

 

 

 

 

 
 

B.

 

Furnace Overheats

 

 

 

 

 

 

 

1. 

The 

HEAT

 light 

goes 

OFF

 with 

the instrument 
switch 

  The controller 

shows a 

very high 

temperature 

  The controller is faulty 

 

 

  The controller 

shows a 

low 

temperature 

  The thermocouple may have been shorted 

out or may have been moved out of the 
heating chamber 
 

 

 

 

 

  The thermocouple may be mounted the 

wrong way round 
 

 

 

 

 

  The controller may be faulty 

 

2. 

The 

HEAT

 light 

does not go off

 

with the 
instrument 
switch 

  The SSR has failed 

“ON” 

  Check for an accidental wiring fault which 

could have overloaded the SSR 

 

Summary of Contents for Carbolite TZF Series

Page 1: ...supply Manuals are supplied separately for the furnace controllers and overtemperature controller when fitted Please read the controller manuals before operating the furnace CONTENTS Section page 1 0 Introduction 2 2 0 Installation 3 3 0 Operation 5 4 0 Error Reference source not found Error Bookmark not defined 5 0 Repairs Replacements 9 6 0 Fault Analysis 12 7 0 Circuit Details 13 8 0 Fuses Powe...

Page 2: ...te that power is being supplied to the elements DANGER of electrical shock read any warning printed by this symbol DANGER hot surface Read any warning printed by this symbol WARNING all surfaces of a furnace may be hot DANGER read any warning printed by this symbol Heat Switch the switch disconnects power to the heating elements unless this switch is off there is a danger of electric shock when in...

Page 3: ...small clip or plate at one end to prevent a wound tube element or a work tube from slipping downwards Mount the furnace so that this clip is at the bottom In models with one thermocouple the thermocouple should be near the top rather than the bottom If the furnace is supplied with a separate work tube or any accessories fit these into position For optimum temperature uniformity insulating plugs sh...

Page 4: ...eferred but not essential Furnace with supply cable either wire directly to an isolator or fit with a line plug Furnace without supply cable a permanent connection to a fused and isolated supply should be made to the internal terminals after temporary removal of the furnace back panel Connection by line plug the plug should be within reach of the operator and should be quickly removable Connection...

Page 5: ...e to display the setpoint press again or hold down to adjust it The display returns to the measured temperature offset when no key is pressed for 0 5 seconds Sometimes a furnace using this type of control does not cool down the end controllers try to remain at the central temperature Should this occur contact Carbolite s technical department 2132 Slave Controller Navigation Diagram B Retransmissio...

Page 6: ...tarts to heat up The Heat light s glow steadily at first and then flash as the furnace approaches the desired temperature or a program setpoint Overtemperature option only If the overtemperature trip operates then an indicator in the over temperature controller flashes and the heating elements are isolated Find and correct the cause before resetting the overtemperature controller according the ins...

Page 7: ...ies to all three controllers Some models covered by this manual are are fitted with 208V elements and are designed for use over the range of voltages 200V 250 the power limit parameter is set accordingly These models may be relocated to a different voltage within the range the power limit should be reset to match the voltage See section 8 2 To check whether the furnace is a 208V model look at the ...

Page 8: ...Engineering Services has a team of Service Engineers capable of repair calibration and preventive maintenance of furnace and oven products at our customers premises throughout the world We also sell spares by mail order A telephone call or fax often enables a fault to be diagnosed and the necessary spare part despatched Each furnace has its own record card at Carbolite In all correspondence please...

Page 9: ... European Association representing the High Temperature Insulation Wool industry www ecfia eu We can provide further information on request Alternatively our service division can quote for any repairs to be carried out at your premises or ours 5 3 Temperature Controller Replacement 201 This controller is fitted to the back of the control panel in many models this can be separated from the base by ...

Page 10: ...ace model its orientation horizontal or vertical and other options and fittings The general procedure is Disconnect the furnace from the electrical supply Remove all outer guards meshes and terminal covers from the furnace body For a horizontally mounted furnace remove the furnace body from its base to reach the bolts or screws which fix the body to the base remove the back panel from the base Dis...

Page 11: ...t in the control system If you have any problems with this procedure please contact our service division 5 7 Fuse Replacement Fuses are marked on the circuit diagram section 7 1 with type codes e g F1 F2 A list of the correct fuses is given in section 8 1 Depending on model and voltage the different fuse types may or may not be fitted If any fuse has failed it is advisable for an electrician to ch...

Page 12: ...r faulty controller There are no lights glowing on the controller Check the supply fuses and any fuses in the furnace control compartment The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault B Furnace Overheats 1 The HEAT light goes OFF with the instrument switch The controller shows a very high temperature The controller is faulty The controller shows a ...

Page 13: ... for 3 phase If type F1 fuse is present one per phase if fitted If type F2 fuse is present one is fitted if type F3 fuse is present one per phase is fitted Note that on 2 or 3 phase models there may be three separate neutrals taken to a common supply terminal depending on EMC filter requirements 7 3 Higher Voltages For 254V or above or 440V 3 phase an isolating transformer is fitted in the control...

Page 14: ...p F2 Auxiliary circuit fuses Fitted on board to some types of EMC filter May be omitted up to 25Amp phase supply rating 2 Amps glass type F On board 20mm x 5mm Other 32mm x 6mm F3 Heat Light fuses May be omitted up to 25 Amp phase supply rating 2 Amps glass type F 32mm x 6mm Customer fuses Required if no supply cable fitted Recommended if cable fitted See rating label for amperage see table below ...

Page 15: ...89 82 75 Custom Models TZF 12 25 550 TZF 12 38 550 TZF 12 75 550 TZF 12 75 900 100 100 89 82 75 Standard Model TZF 12 100 900 from April 2000 100 100 89 82 75 89 82 75 Other possible exceptions Further custom furnaces based on the standard models may be designed with power limits other than 100 Furnaces supplied for use on 254 Volt typically have a power setting of 89 of the limit for the correspo...

Page 16: ...Covered by this Manual MODEL Max Temp Max Power Work Tube Bore Work Tube Length Heated Length Uniform Zone Length1 Net Weight2 C kW mm mm mm mm kg Tube furnaces with a ceramic work tube wound with resistance wire TZF 12 38 400 1200 C 1 5 38 450 400 305 15 TZF 12 65 550 1200 C 2 0 65 600 550 392 25 TZF 12 75 700 1200 C 3 0 75 750 700 510 28 TZF 12 100 900 1200 C 4 5 100 950 900 745 35 also models b...

Reviews: