3.0 301 Controller
3.12
Glossary of Terms
Setpoint (SP)
The target temperature the furnace or oven is trying
to reach.
°C
Setpoint Ramp
Rate (SPrr)
The speed at which the furnace or oven is allowed to
heat up or cool down
°C/ Min
Element
The heating device used in the furnace or oven.
-
Thermocouple
A thermoelectric device for measuring temperature.
-
PID
Proportional Integral Derivative: the control system
used by the controller
-
Over-
Temperature
The condition which a furnace or oven may enter if
part of the main control circuit fails.
-
Over-
Temperature
Protection
A system to prevent the product or process being
damaged if it has gone into an Over-Temperature
condition.
-
Furnace or Oven
This refers to the product purchased from Carbolite
Gero
-
3.13
Controller Fault
Fault Code Diagnostic Table
Error
Code
Explanation
Actions
S.br
Temperature
sensor
failure
Check all terminal connections between the temperature
sensor (thermocouple) and temperature controller. It is
recommended to loosen then tighten the screws in the
terminal blocks in case the connections are oxidised. If this
does not correct the error then replace the furnace or oven
temperature sensor (thermocouple).
""""
Input over
range
-ooo
Input under
range
E
-
followed
by
numerical
code
Controller
Error
Turn the furnace or oven off and back on to see if this clears
the error. If not contact Carbolite Gero Service – (see back
cover)
38