PASSWORD
The password protects the modification of the
Dosing
,
Utility
and
Relay
menus. It does not block
Calibration
and
Set
up
. By default the password is not present and it is therefore possible to access all the programming.
First definition of the password:
In the Set up menu, select "Password" by the cursor and
press
ENTER
.
Enter the password digit by digit selecting the number by ▼
▲ button and pressing
ENTER
.
After the selection of the 4th digit, press
ENTER
to save the
password.
Take note of it carefully to not forget it. If it is not
possible to remember the password call the Customer
Service Assistance.
To esc programming and return to measurement, press the
ESC
button for 3 seconds.
Change password:
In the
Set up
menu, select "Password" by the cursor and
press
ENTER
.
Enter the old password digit by digit selecting the digit by ▼
▲ buttons and press
ENTER
.
After entering the 4th digit, pressing
ENTER
it is possible to
define the new password.
Enter the new password digit by digit selecting the digit by ▼
▲ buttons and press
ENTER
.
After entering the 4th digit, press
ENTER
to save the new
password.
Take note of it carefully to not forget it.
If it is not possible to
remember it, call the Customer Service Assistance.
If you want to esc the programming and return to
measurement, press the
ESC
button for 3 seconds.
ENABLE (OPTIONAL)
The enable input (230Vac-110Vac) gives to the pump the consent to dose. In swimming pool plants this input is
connected in parallel with the recirculation pump power supply so that if the filtration is stopped the pump can not dose. In
order not to damage the electronic board, do not connect the enable directly in parallel with the recirculation pump power
supply, but always use a contactor / relay (see installation manual). In the case that your pump has the enable option (see
position B page 19: CONNECTIONS), you can choose to activate this input or not as follows:
In the
Set up
menu, select "Enable" by the cursor and
press
ENTER
.
Use the ▼ ▲ buttons to select whether to enable (“active”)
or disable (deactive) this input and press
ENTER
.
If you want to esc programming and return to
measurement, press
ESC
for 3 seconds.
If your pump has not included the enable option then it will not be possible to access the “Enable” submenu and the
flashing message "DEA" will appear: