ELECTROMAGNETIC COMPATIBILITY
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Emmission Test
Compliance
Electromagnetic Environment Guidance
Group 1
Class B
Class A
Complies
This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
This device is suitable for use in all establishments, including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations
on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
line(s) to line(s):
±1 kV. line(s) to earth:
±2 kV. 100 kHz repetition frequency
5% UT (>95% dip in UT)
for 0.5 cycle 40% UT (60% dip in UT)
for 5 cycle 70% UT (30% dip in UT)
for 25 cycle <5% UT (>95% dip in UT)
for 5s
3 A/m
50Hz/60Hz
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
5% UT (>95% dip in UT)
for 0.5 cycle 40% UT (60% dip in UT)
for 5 cycle 70% UT (30% dip in UT)
for 25 cycle <5% UT (>95% dip in UT)
for 5s
3 A/m
50Hz/60Hz
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of this device requires continued operation during
power mains interruptions, it is recommended that this device be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electromagnetic Environment Guidance