background image

13

14

Table 204: Guidance and manufacturer’s declaration –

electromagnetic immunity

The VeriChip H2 Reader is intended for use in the electromagnetic

environment specified below.  The customer or the user of the

VeriChip H2 Reader should assure that it is used in such an environment.

Immunity

Test

IEC 60601

test level

Compliance

level

Electromagnetic

environment – guidance

Portable and mobile RF

communications equipment

should be used no closer to

any part of the VeriChip H2

Reader, including cables,

than the recommended

separation distance

calculated from the

equation applicable to the

frequency of the transmitter.

Immunity

Test

IEC 60601

test level

Compliance

level

Recommended

separation distance

Conducted 

RF IEC 

61000-4-6

Not

applicable

Not

applicable

Not applicable since the

VeriChip H2 Reader is not

connected by cabling and

does not use a power cord.

Radiated

RF IEC 

61000-4-3

3 V/m 

80 MHz to 

2.5 GHz

3 V/m

80 MHz to 800 MHz 

d

= 1.2 

P

800 MHz to 2.5 GHz 

d

= 2.3 

P

Where 

P

is the maximum

output power rating of the

transmitter in watts (W)

according to the transmitter

manufacturer and 

d

is the

recommended separation

distance in meters (m).

Table 204: Guidance and manufacturer’s declaration –

electromagnetic immunity (continued on next page)

Table 204: Guidance and manufacturer’s declaration –

electromagnetic immunity (continuation)

Immunity

Test

IEC 60601

test level

Compliance

level

Recommended

separation distance

Radiated RF 

IEC 61000-4-3

3 V/m 

80 MHz to

2.5 GHz

3 V/m

Field strengths from

fixed RF transmitters,

as determined by an

electromagnetic site

survey, 

a

should be less

than the compliance

level in each frequency

range.  Interference

may occur in the

vicinity of equipment

marked with the

following symbol:

NOTE 1:  At 80 MHz and 800 MHz, the higher frequency range

applies.

NOTE 2:  These guidelines may not apply in all situations.

Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for

radio (cellular/cordless) telephones and land mobile radios, amateur

radio, AM and FM radio broadcast and TV broadcast cannot be

predicted theoretically with accuracy.  To assess the electromagnetic

environment due to fixed RF transmitters, an electromagnetic site

survey should be considered.  If the measured field strength in the

location in which the VeriChip H2 Reader is used exceeds the

applicable RF compliance level above, the VeriChip H2 Reader 

should be observed to verify normal operation.  If abnormal

performance is observed, additional measures may be necessary,

such as reorienting or relocating the VeriChip H2 Reader.

Summary of Contents for VeriChip H2

Page 1: ...O P E R A T O R S M A N U A L AN APPLIED DIGITAL COMPANY...

Page 2: ...Press Button F1 and release to power on the Reader 2 The Reader will enter a self diagnostic mode displaying the following on the LCD screen a Initializing b VeriChip H2 Reader BT Master Rev x xx c A...

Page 3: ...ally enter the ID number Turning the Reader Off The Reader can be turned off manually at any time by simultaneously pressing and holding down the F2 and F3 buttons If the Authorization Code is enabled...

Page 4: ...al should be set to 9600 Baud 8 data bits one stop bit no parity 8 N 1 To enter the configuration mode the user should send a predetermined command from the host to VH2R letter c in lower case The pow...

Page 5: ...tion in USB Mode To operate in this mode the user should perform the following actions Connect the USB cable from the host computer to the VH2R Reader The USB port is located in the rear of the case L...

Page 6: ...l not require entering the passkey unless the devices become unpaired The operation is performed as follows Tooperateinthismodetheusershouldperformthefollowingactions Bring the VH2R and host system wi...

Page 7: ...the electromagnetic environment specified below The customer or the user of the VeriChip H2 Reader should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance le...

Page 8: ...ty continued on next page Table 204 Guidance and manufacturer s declaration electromagnetic immunity continuation Immunity Test IEC 60601 test level Compliance level Recommended separation distance Ra...

Page 9: ...ing the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and...

Page 10: ...in which it will be used Warning The VeriChip H2 Reader generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful...

Page 11: ...measures Reorient or relocate the receiving antenna Increasetheseparationbetweentheequipmentandthereceiver Consultthedealeroranexperiencedradio TVtechnicianforhelp Modifications Any changes or modifi...

Reviews: