13
14
Table 204: Guidance and manufacturer’s declaration –
electromagnetic immunity
The VeriChip H2 Reader is intended for use in the electromagnetic
environment specified below. The customer or the user of the
VeriChip H2 Reader should assure that it is used in such an environment.
Immunity
Test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the VeriChip H2
Reader, including cables,
than the recommended
separation distance
calculated from the
equation applicable to the
frequency of the transmitter.
Immunity
Test
IEC 60601
test level
Compliance
level
Recommended
separation distance
Conducted
RF IEC
61000-4-6
Not
applicable
Not
applicable
Not applicable since the
VeriChip H2 Reader is not
connected by cabling and
does not use a power cord.
Radiated
RF IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to 800 MHz
d
= 1.2
√
P
800 MHz to 2.5 GHz
d
= 2.3
√
P
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Table 204: Guidance and manufacturer’s declaration –
electromagnetic immunity (continued on next page)
Table 204: Guidance and manufacturer’s declaration –
electromagnetic immunity (continuation)
Immunity
Test
IEC 60601
test level
Compliance
level
Recommended
separation distance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey,
a
should be less
than the compliance
level in each frequency
range. Interference
may occur in the
vicinity of equipment
marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which the VeriChip H2 Reader is used exceeds the
applicable RF compliance level above, the VeriChip H2 Reader
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary,
such as reorienting or relocating the VeriChip H2 Reader.