10. Safety instructions
Ensure there are no objects blocking the
regular usage of the wheels of the
rollator.
Risk of burns – Be careful when driving in
hot or cold environments (sunshine,
extreme cold, saunas, etc.) for a
sufficient amount of time and when
touching.
Before every usage, check that the star
buttons for your handgrips are firmly
locked.
Verify if the safety against unwanted
folding/unfolding is functioning properly.
Only proceed moving with your rollator
when you are standing comfortably and
when both your hands are resting on the
pushing handgrips.
Do not hold the pushing handgrips with
wet hands. If you do so, you can loose
your grip and loose your balance.
When you detect that your body is getting
tired, stop your rollator by placing the tips
on the ground and have a seat on the
bench.
The maximum users weight for your
rollator is 130 kg
and can only be used
on a flat underground.
Do not use your rollator on a staircase or
bumps higher than 40 mm.
Avoid roads with a very steep inclination.
Avoid wet floors.
Do not use your rollator to transport
people or items.
Take note of the instructions for care and
service. The manufacturer is not liable for
damage caused by improper
servicing/care.
11. Care
It is recommended to use a soapy solution for
cleaning. The cleaning liquid should in any
case have a pH value of at least 6. Do not
use a cleaning agent which contains solvents.
Observe the instructions of the cleaning liquid
used. Use a damp cloth and lukewarm water
for cleaning. Avoid soaking. Do not use metal
brushes or any other sharp-edged or pointed
cleaning utensils. In the case of damage to
the varnish or the woodwork, consult your
specialist dealer. Regular care would ensure
that your rollator is preserved in a perfectly
functional condition.
12. Disinfection
For sterilisation use only disinfectants suitable
for treating varnished metal and wood
surfaces. Observe the instructions of the
cleaning liquid used. Use only proved
processes (for scrubbing disinfection) and
disinfectants included in the list of the Robert
Koch Institute (information can be obtained
from: www.rki.de). Disinfections may only be
undertaken by trained persons. Consult your
specialist dealer.
13. Repairs / Service / Inspection
Lifetime of the rollator is influenced by its use,
storage, regular maintenance, servicing and
cleaning.
You should use only genuine VERMEIREN
replacement parts. Repairs and renovations
may only be undertaken by trained persons.
Make use of the services offered by your
specialist dealer. He would gladly assist you
further in regard to service and repairs.
Before you resume use of the rollator, it should
be serviced according to the following criteria:
Completeness
Folding
mechanism
Check the castors (stability, tracking, fixing
screws)
Frames (deformation, stability,
connections)
Pushing handgrips (safe locking from
height adjustment system, good support
from the handgrip)
Seat, shopping basket on wear or damage
Tips on wear or damage and replace if
necessary. To avoid damages on floors.
Cleaning
Damaged varnish (remove possible rust
and repaint)
Every 8 weeks:
Lubricating (moving parts shall be
greased)
Every 6 months or for every new user, it
should be serviced according to the
following criteria:
General
review
Disinfection