Albatros II
2022-08
Maintenance
22
EN
NL
5.2
Shipping and storage
Shipping and storage should happen according to the technical specifications in §7. Make
sure that the patient hoist is stored dry.
Prevent cables from being crushed or bent with kinks.
Disconnect the patient hoist from the socket.
Provide sufficient covering or packaging to protect the patient hoist against rust and foreign
bodies. (e.g. salt water, sea air, sand, dust).
Store all removed parts together in one place (or mark them if necessary) to avoid mixing
up with other products when re-assembling (e.g. the charger).
Make sure no objects are stored on or against the patient hoist and its components to
prevent damage.
5.3
Cleaning
Risk of damage by moisture
Never use a hose or high-pressure cleaner.
Use a mild commercial detergent to remove stubborn dirt.
Stains can be removed by using a sponge or a soft brush.
Do not use strong cleaning liquids like solvents, nor use hard brushes.
5.3.1 Plastic parts
Clean plastic parts of your patient hoist with commercial plastic cleaners. Read the specific
product information and only use a soft brush or soft sponge.
5.3.2 Coating
The patient hoist consists of steel components. The high quality of the surface layer guarantees
optimal protection against corrosion. If the outer coat is damaged by scratching or in some other
way, get your specialist dealer to repair the affected surface.
When cleaning, only use warm water and normal household detergents and soft brushes and
cloths. Ensure that no wetness gets into the tubes.
5.3.3 Electronics casing
L
WARNING: Risk of injury - The emergency stop must be activated before maintenance
work, so that no unintentional adjustment movements can be made.
Wipe the control box, motor casing and manual controls down with a cloth moistened by a few
drops of a commercial domestic cleaner. Do not use abrasives or sharp-edged polishing
equipment like a metal scrubber or brush, as these can scratch the surface of the manual control
and destroy its water repelling property.
Regularly check the plug connectors for corrosion or damage, as these could affect the functional
integrity of the electronics.
The manufacturer will not accept liability for damages caused by insufficient maintenance.
5.4
Disinfection
All parts of the patient hoist can be treated by scrubbing with a commercial disinfectant. Follow
the instructions on the disinfectant.
Washing instructions for slings can be found in the slings manual and on the label of each sling.
CAUTION
Summary of Contents for Albatros
Page 4: ...This page is intentionally left blank ...
Page 32: ......
Page 88: ......
Page 114: ...Albatros II 2022 08 Technische Daten 26 EN NL DE 7 Technische Daten Vorwärtsrichtung ...
Page 144: ...Albatros II 2022 08 Specifiche tecniche 28 EN NL IT 7 Specifiche tecniche direzione avanti ...
Page 176: ......
Page 203: ...Albatros II 2022 08 Parametry techniczne 27 EN NL PL 7 Parametry techniczne Do przodu ...
Page 206: ......
Page 232: ...Albatros II 2022 07 Technické údaje 26 EN NL PL CS 7 Technické údaje Směr dopředu ...