E
SQUOD_SU
06/2009
24
CONFORMITY
The SQUOD SU electric wheelchair complies with the requirements of the European directive:
- 93/42/EEC (Medical Products Directive)
and the product norms:
- EN 12182: 1999
- (DIN) EN 12184: 1999
DISPOSAL
The manufacturer is responsible for taking back and recycling the electric wheelchair and shall
meet the requirements of European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment. You can find out from your local waste disposal company where you can take the
electrical wheelchair to be recycled free of charge. It may not be disposed of with domestic
waste.
Your dealer will also be able to advise you should you have any queries.
SHARK 2 DISPLAYS
Display Unit
Description
Meaning
Remarks
All LEDs dark.
System off.
All LEDs are lit.
System on.
LEDs show the charge
level.
Left red LED
flickering.
Low battery level.
Charge battery
immediately.
Progress indicator
from
“right to left” /
"green to red"
Shark system blocked.
To unlock, press the horn
button twice within 10
seconds after switching the
steering unit on.
Progress indicator from
“left to right” /
"red to green"
with subsequent
charge level display.
Shark is being
programmed, is connected
to the charger and/or is
busy charging.
The flickering LEDs
indicate the present
loading status.
All LEDs flickering
slowly.
After starting the system,
the joystick is not in the
neutral position.
Put the joystick in the
neutral position (let it go).
ERROR CODES
Symbol flickering fast at intervals. The number of flashes per interval provides the information below
(the following actions and checks should only be carried out by authorised individuals).
Number
Problem/Fault
Checks
0
There is no indication of
the charge level of the
batteries after the steering
unit is switched on.
1. Check whether the battery plug is inserted fully and correctly
connected to the socket of the steering unit.
2. Check whether the two batteries are connected correctly.
3. Check whether the batteries are charged.
4. Check whether the fuses of the steering unit and the batteries are
defective or have burnt out.
1
Operating error
1. Possible safety shut down or operating error (joystick)
2. Put the joystick in neutral position and restart the system.
2
Battery error
1. Check batteries and cabling; if necessary charge.
2. If necessary, replace batteries.
Progre
ss
indicat
or
Progress
indicator
charge
level