Vernacare Compact Plus Technical Manual-
Revision 2-Issued by N.Shields 14.2.2020
Page | 17
4.5.
Electrical
4.5.1.
Single-phase electrical system (230v, 1 Ph, 50 HZ) (110v, 1 ph, 60
HZ)
The machine is supplied with a 3 metre length of 1.5mm flex to BS6500 which
shall be connected to:
• 230 Volt Machine – 10 Amp MCB to BS EN60898 Type C or fused 10 Amp to
IEC 60269.
• 110 Volt Machine – 16 Amp MCB to BS EN60898 Type C or fused 16 Amp to
IEC 60269.
Notes:
1.
All electrical installation to comply with current I.E.E. regulations.
2.
Mains supplies to be protected by:
1 Phase 230v ......... 10 Amp approved I.E.C. fuse.
1 Phase 110v…….. 16 Amp approved I.E.C. fuse.
3.
For all installations, an I.E.E. approved disconnection device
must
be fitted to the final
installation. This device shall be positioned in close proximity to the Compact Plus and
within easy reach of the operator & shall also be clearly marked as being the
disconnecting device for the machine.
4.5.2. Back-up generator tests / mains electrical supply surges.
Damage to internal electrical components of the Compact Plus macerator can occur as
a result of voltage surges / spikes occurring in the electrical supply to the machine. The
likelihood of such damage occurring is increased when frequent testing of back-up
electrical supplies are carried out.
To prevent such damage, Vernacare recommends that transient voltage suppressors
(in accordance with IEC class 3 surge protection) be installed in the electrical supply to
the machine.
4.6.
Environment
Indoor use only. Altitudes up to 2000m. Temperatures 5
C to 40
C.
Maximum relative humidity 80% for temperatures up to 31
C, decreasing linearly to 50%
relative humidity at 40
C.