Vestfrost Solutions 21
EN
Your device is equipped with a remote alarm
connection to send alarm signals to a building
management system, a monitoring unit, visual
or acoustical warning device etc�
The alarm output is managed freely by the
built in controller of the unit�
Alarm criteria may be too low temperature,
too high temperature, power failure, open
door or similar�
The alarm output is an electrical floating
output (dry contact) which acts like a ON/OFF
switch�
A SPDT (single pole, dual throw) relay with
maximum switch rating of 250V 3A is main-
tained by the build-in XW737K controller unit�
The relay terminals are explained this way:
COM: COMMON terminal
NO: NORMALLY OPEN terminal�
Connects to COM when the alarm is activated
NC: NORMALLY CLOSED terminal�
Disconnects from COM when the alarm is
activated
An alarm interface cable of maximum
30 meters can be attached without the use of
tools to the 3-pole alarm clamp connector�
Wire thickness must be within 0�5mm2 to
2�5mm2 and fitted with core crimp terminals�
Ala
rm
XW
73
7K
16
17
18
3. COM
2. NO
1. NC
BK
OR
WH
Alarm
Fig� Internal XW737K controller with
build-in alarm relay
Remote alarm function
Summary of Contents for ULTF 40
Page 2: ......
Page 33: ......
Page 34: ...2 Vestfrostsolutions com DE...
Page 64: ...2 Vestfrostsolutions com FR...
Page 94: ...32 Vestfrostsolutions com FR...
Page 95: ...Vestfrost Solutions 33 FR...
Page 96: ...34 Vestfrostsolutions com FR...
Page 97: ...Vestfrost Solutions 35 FR...
Page 98: ...36 Vestfrostsolutions com FR...
Page 99: ...Vestfrost Solutions 37 FR...