Recharching refrigerant from a service bottle
The recharging shall take place in a room
where the ambient temperature is be-
tween 20 and 25 °C.
1. Connect service bottle, compressor
and vacuum pump (A) to a T
2. Open valve (C) and let vacuum pump
run for minimum 30 min
3. Close valve (C)
4. Place the bottle on a scale if possi-
ble.
5. Open valve (B) slowly
6. When the refrigerant stops flowing
into the system, start the compres-
sor.
7. Let the compressor run for 5 min
8. Check amount of refrigerant charged
on the scale, should be approx. the
same as on the rating plate
9. Repeat from 5 if necessary
Note! If you start the compressor at point
8 with the valve open then you can ruin
the compressor.