36
List of parameters VTS258:
Group
Parameter
Description
Original
Vis. Level
Min.
Maxi.
Unit
Clock and recording setting
itP
Recording interval
5
Pr1
1
255
min
Clock and recording setting
rC1
First probe recording enable
YES
Pr2
Clock and recording setting
rC2
Second probe recording
NO
Pr2
Clock and recording setting
rC3
Third probe recording enable
NO
Pr2
Clock and recording setting
rC4
Fourth probe recording enable
YES
Pr1
Clock and recording setting
rCb
Start recording SET key enabling
YES
Pr1
Clock and recording setting
EU
Date format
EU
Pr1
Clock and recording setting
rSd
Data erase
NO
Pr2
Clock and recording setting
rSA
Alarms erase
NO
Pr2
Regulation
Set
Set point
-87
Pr1
-95
-55
°C
Regulation
Hy
Differential
1
Pr2
1
26
°C
Regulation
LS
Minimum set point limit
-95
Pr2
-100
-87
°C
Regulation
US
Maximum set point limit
-55
Pr2
-87
150
°C
Probe inputs
ot
Regulation probe calibration (term. 1-2)
-1
Pr2
-12
12
°C
Probe inputs
P2P
Evaporator probe presence (term. 2-3)
NO
Pr2
Probe inputs
oE
Evaporator probe calibration
0
Pr2
-12
12
°C
Probe inputs
P3P
Third probe presence (term. 4-5)
NO
Pr2
Probe inputs
o3
Third probe calibration
0
Pr2
-12
12
°C
Probe inputs
P4P
Fourth probe presence (term. 5-6)
NO
Pr2
Probe inputs
O4
Fourth probe calibration
0
Pr2
-12
12
°C
Probe inputs
odS
Outputs activation delay at start up
0
Pr2
0
255
min
Probe inputs
AC
Anti-short cycle delay
5
Pr2
0
30
min
Probe inputs
Con
Compressor ON time with faulty probe
60
Pr2
0
255
min
Probe inputs
COF
Compressor OFF time with faulty probe
5
Pr2
0
255
min
Display
CF
Temperature measurement unit
°C
Pr1
Display
rES
Resolution (for °C)
in
Pr1
Display
rEd
Remote display
P1
Pr2
Display
dLy
Display delay
0.00
Pr2
min
Defrost
dtE
Defrost termination temperature
4
Pr2
-100
150
°C
Defrost
ldF
Interval between defrosts
1
Pr2
1
120
hour
Defrost
MdF
(Maximum) duration of defrost
0
Pr2
0
255
min
Defrost
dFd
Display during defrost
rt
Pr2
Defrost
dAd
Defrost display time ou
t
0
Pr2
0
255
min