VC series bench scales Technical Manual
- 13 -
If display will be shown any error or incorrect value, repeat the procedure again.
7.2. Linear Calibration
The linearity deviation caused by the performance of the weighing unit. The digital
linearization function can reduce the linearity deviation using weighing points during
the zero and capacity. Up to three weighing points can be specified.
Enter the function by pressing
, display will be shown
Make sure there are no loads on the platform and wait few seconds for stable
indicator on.
Load the first calibration mass weight on the platform (mass weight should
be1/4 of the max capacity) and wait few seconds for to show display
Load the second calibration mass weight on the platform (mass weight should
be2/4 of the max capacity) and wait few seconds for to show display.
Load the third calibration mass weight on the platform (mass weight should
be3/4 of the max capacity) and wait few seconds for to show display
Load the fourth calibration mass weight on the platform (mass weight should
be4/4 of the max capacity) and wait few seconds for to show display
After the calibration the display will start a self test. Remove the load from platform
during the test. Display will come to weighing mode automatically.
If display will be shown any error or incorrect value, repeat the procedure again.
linear
Load 0
Load 1
Load 2
Load 3
Load 4
pass