17.1.1. Manual internal calibration
Procedure:
x
While in weighing mode press
,
x
The scale program starts to check stability conditions for the
calibration process and displays the following message:
x
Then the program automatically goes to the internal calibration
procedure which is signalled by the following message:
x
After completion of the calibration process program returns
to the weighing mode,
x
Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
Notice:
1. It should be remembered that internal calibration should
be performed with unloaded pan with keeping possibly
constant ambient conditions.
2. If the calibration process lasts longer than 15 seconds scale
software will react with
<Err8>
displayed and a short sound
and then the calibration procedure will start again.