- 66 -
SUI_I CR LF
- command comprehended but cannot be executed
SUI_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in current unit is returned immediately
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
Example:
S U I CR LF
– computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command executed and mass
returned
20.4.7. Switch on continuous transmission in basic unit
Syntax:
C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
space
sign
mass
space
unit
CR
LF
20.4.8. Switch off continuous transmission in basic unit
Syntax:
C0 CR LF
Possible answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed