Electromagnetic Compatibility
The device meets the requirements of EN 60601-1-2. All the accessories
also meet the requirements of EN 60601-1-2 when in use with this device.
Warnings and Cautionary Advices
Using accessories other than those specified in this manual may
result in increased electromagnetic emission or decreased
electromagnetic immunity of the equipment.
The device or its components should not be used adjacent to or
stacked with other equipment.
The device needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information
provided below.
Other devices may interfere with this device even though they meet
the requirements of CISPR.
When the inputted signal is below the minimum amplitude provided
in technical specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the
performance of this device.
Other devices that have RF transmitter or source may affect this
device (e.g. cell phones, PDAs, and PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The Health Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC61000-3-2
Class A
Voltage Fluctuations / Flicker
Emissions IEC 61000-3-3
Complies