34
Parameter item
Specifications
Safety features
Electric shock
protection type
Class II device
Electric shock
protection level
Type BF
Operating mode
continuously working
Waterproof protection
level
IP22
Safety to flammable
anesthetic gas
Can not be used in the presence of flammable
anesthetic gas mixed with air or flammable
anesthetic gas mixed with oxygen or nitrous
oxide
Installation and use
classification
Portable device
Power connection
Adapter detachable power cord
Pressure
Setting range
4.0-25cmH
2
O (EPAP)
Maximum ultimate
pressure
40cmH
2
O(Single failure
)
25cmH
2
O
(
Normal status
)
Static pressure control
accuracy
±0.5 cmH
2
O
Dynamic pressure
control accuracy
±1 cmH
2
O
Flow
Maximum flow rate
(1/3, 2/3 and
maximum value of
the maximum
adjustable pressure)
8 .5cmH
2
O:≥120 L/min
;
17 cmH
2
O
:
≥120L/min
25 cmH
2
O
:
≥120 L/min
Tidal volume
(BTPS)
Tidal volume
measurement range
0-3000 mL:±(20% of actual reading)
Other scope: no definition
Tidal volume control
±(50 mL+20% of setting value)