15
9. Electromagnetic Compatibility
The device meets the requirements of EN 60601-1-2. All the accessories also meet the requirements
of EN 60601-1-2 when in use with this device.
Warnings and Cautionary Advices
Using accessories other than those specified in this manual may result in increased
electromagnetic emission or decreased electromagnetic immunity of the equipment.
The device or its components should not be used adjacent to or stacked with other equipment.
The device needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided below.
Other devices may interfere with this device even though they meet the requirements of
CISPR.
When the inputted signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of this device.
Other devices that have RF transmitter or source may affect this device (e.g. cell phones, PDAs,
and PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations /
Flicker Emissions IEC
61000-3-3
Complies
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the Health Monitor should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic
environment - guidance