5.15 Spectrum Analyzer
The CX300’s Spectrum Analyzer is a full-featured spectrum analyzer which utilizes the RF
Input or RF Duplex connector. The span is user selectable, as is a wide range of resolution
bandwidth (RBW) values. The RBW filters are realized entirely in the CX300’s digital IF,
providing ideal bandwidth accuracy and shape factor. Video bandwidth (VBW) and sweep
time settings can be auto-coupled or manually adjusted for maximum flexibility and ease
of use.
The Spectrum Analyzer supports a variety of detectors which allows for accurate power
measurements of wide-band and noise-like signals. Up to six traces can be displayed on
the screen, each specifiable as clear/write, max hold, min hold, average, blank, or view.
The number of trace points and number of trace averages are also user-configurable.
Marker functionality includes peak and next peak functions, with specified level thresholds,
delta markers, and frequency counter and noise markers. The Spectrum Analyzer also
provides zero-span functionality, allowing time domain analysis.
5.15.1 Spectrum Analyzer Controls and Settings
The following controls and settings are used to configure the Spectrum Analyzer:
5.15.1.1 Frequency and Span Controls and Settings
Frequency and span settings define the frequency range used to evaluate signals. If you
know the frequency of the signal being evaluated, the center frequency should be set to
match the signal’s frequency. If you are investigating signals that are within a particular
frequency range, it is best to enter a start and stop frequency to define the span.
Table 5-36 Spectrum Analyzer Frequency and Span Controls and Settings
Control/Setting
Description
Center Freq
This field sets the horizontal center of the display to a specific
frequency. The left and right sides of the graticule correspond
to the start and stop frequency values.
Span
This setting defines the frequency range around the center
frequency. The span frequency readout describes the total
displayed frequency range. To determine span frequency per
horizontal graticule division, divide the frequency span by 10.
Start Freq
This field sets the lower end of the frequency span which is the
lower rance frequency starting at the left side of the plot.
Stop Freq
This field sets the upper end of the frequency span which is the
upper range frequency on the right side of the plot.
Full Span
Pressing this button sets the span to full span showing the full
frequency range of the instrument.