FVAi-2030 Benchtop Microscope User Guide, 22112369-339 R000, Standard
2-4
Getting started
Mounting fiber adapters and connectors
Mounting fiber adapters and connectors
Important:
Store fiber adapters in dust-proof containers while not installed on a microscope.
Step 1 If present, unscrew the lens cap or fiber adapter from the adapter mount, and store it in a dust-proof
container.
Step 2 Place the required fiber adapter onto the mount, and finger-tighten the adapter to secure it. Do not
over-tighten the adapter.
Step 3 Power on the microscope, and access Live view on the LCD touchscreen (see
).
Step 4 Clean the fiber connector to be inspected.
Step 5 Ensure that the fiber connector is correctly oriented, and then insert it into the fiber adapter.
Step 6 Ensure that the fiber end face is properly centered within the view.
Note:
The message
Auto Focus Error: Unable to locate the fiber
will appear when you
attempt to use the
Autofocus
tool while the fiber end face is poorly centered.