FVAi-2030 Benchtop Microscope User Guide, 22112369-339 R000, Standard
2-5
Getting started
Mounting fiber adapters and connectors
Step 7 If required, use the M5 hex screwdriver supplied with the microscope to adjust the vertical and
horizontal position of the lens.
Note:
You can use the Fiber Centering Limit setting and marker as a reference for the adjustment. For
information, see
“Managing the Fiber Centering Limit setting and marker” on page 2-16
.
Insert the end of the M5 hex screwdriver into an adjustment pilot hole until it engages, and then
carefully turn the screwdriver in either direction, referring to the touchscreen as the position of the lens
changes.
— End —
CAUTION
• Use only the M5 hex screwdriver supplied with the microscope.
• Do not force the screwdriver beyond the adjustment limits.