FVAi-2030 Benchtop Microscope User Guide, 22112369-339 R000, Standard
4-9
Using the FVAi-2030 microscope with FiberChekPRO
Creating analysis profiles for the FVAi-2030 microscope
Step 6 For each available zone, select the tab for the zone and specify settings as required:
▪
Name
,
Inner Diameter,
Outer Diameter
, and
Outer Diameter Color
for the zone.
▪ Scratches:
Specific acceptance criteria for scratches.
▪ Defects:
Specific acceptance criteria for defects, including any non-scratch defect on the connector
(dirt, debris, pits, chips). Specific criteria can be defined for both
Individual
and
Combined
defects.
Select
Enable Advanced Criteria
to define additional options to fine-tune acceptance criteria that
address specific requirements.
▪
Exclusion Zone:
Inner and outer radius in which defects and scratches are not counted toward the
acceptance criteria.
Step 7 Select
OK
to close the dialog for the profile.
A new FVAi-2030 profile is automatically added to the
On Device
list. An FVAi-2030 profile created using
a copy of an existing profile is automatically added to the same list from which the original profile was
selected (
).
Step 8 Select
OK
in the navigation pane to return to the Inspection window.
— End —