FVAi-2030 Benchtop Microscope User Guide, 22112369-339 R000, Standard
2-10
Getting started
Applying a profile and an optical setting
Applying a profile and an optical setting
The FVAi-2030 microscope provides factory-set analysis profiles and optical settings for use in Pass/Fail testing
(
Table 2-3
). A profile defines Pass/Fail thresholds for a fiber type, while an optical setting defines the lighting and
sizing calibration applied during a test.
Refer to
Table 2-3
to ensure that the appropriate profile and optical setting for the fiber connector are applied
before performing a test during standalone operation of the microscope.
Note:
FiberChekPRO provides tools that let you create customized profiles (see
“Creating analysis profiles for
the FVAi-2030 microscope” on page 4-6
) and optical settings. You can also manage factory-set and custom
profiles and optical settings with FiberChekPRO, making them available on or removing them from the
microscope as required. For information, including information about applying profiles and optical
settings using FiberChekPRO, see the
FiberChekPRO Automated Fiber Inspection and Analysis Software
User Manual
.
Step 1 In Live view, select
Inspection Settings
, and then do either of the following:
▪ Select
PROFILE
to access the profiles available on the microscope, select a profile, and then select
Back
to return to the
Inspection Settings
menu.
Table 2-3: Factory-set analysis profiles and optical settings
With Analysis Profile...
Use Optical Setting...
FVAi-2030 SM 1.25 mm UPC (Polish detect)
FMAX-LC-SL35
FVAi-2030 SM 1.25 mm APC (Polish detect)
FMAX-LCA-SL35
FVAi-2030 SM 2.5 mm UPC (Polish detect)
FMAX-SC-SL30
FVAi-2030 SM 2.5 mm APC (Polish detect)
FMAX-SCA-SL30
FVAi-2030 MM50 1.25 mm UPC (Polish detect)
FMAX-LC-SL35
FVAi-2030 MM50 2.5 mm UPC (Polish detect)