MPOLx-84/85 (P)
47
6 L
OSS
/L
ENGTH
T
EST
O
PERATION
D
EFINING
A
TEST
CONFIGURATION
3.
Tap the
[
Negative Loss Warning
]
button to activate or
deactivate the warning.
To activate or deactivate the unsaved result warning:
I
The
Loss/Length
application is selected.
1.
Press the
[
A
]
key.
2.
Tap the
[
More
]
button.
3.
Tap the
[
Unsaved Result Warning
]
button to activate or
deactivate the warning.
Defining a test configuration
In
Loss/Length
mode a test configuration allows you to define
various settings like the test reference method or the cable and
connector settings. When performing a test a predefined test
configuration can be recalled and set to active to assign it.
As factory setting one test configuration named
default
is
available, which can be edited but not deleted.
After deleting the test configuration
default
, a new (empty) test
configuration called
default
will be created automatically.
In the
[
Test Configuration
]
menu, new configurations can be
created or existing configurations can be copied, edited, or
deleted.
NOTE:
Test configurations can only be created, deleted, or edited with
the MPOLP-85/MPOLP-85P. The MPOLS-85/MPOLS-85P only
allow displaying the settings of the active test configuration.
Creating a new test configuration
I
The instrument is in
Loss/Length
mode.
J
Press the
[
A
]
key and tap the
[
Test Configuration
]
button.
The available test confgurations are displayed.