34
3-6-2
Comparator function setting
-
-
When to use HI/OK/LO buzzer, be sure to set <19 BUZZER> to ON.
(Refer to 3-12 Buzzer setting.)
Lower limit, reference and upper limit can be also set by [1](LOW) key, [2](OK) key and [3](HIGH)
key at “comparator indication” display.
1
Select the comparator function.
- Press [Menu] key, then press [1]-[4] keys to
go to <131 ACTIVATE>
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
OFF: OFF
H / L: Upper and lower limits valid
HIGH: Upper limit alone valid
LOW: Lower limit alone valid
- Press [4] key to fix.
2
Select the discriminant condition.
- Press [1]-[2] keys to go to <132 CONDITION>
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
FULL: At all times
STBL: Only at stable times
- Press [4] key to fix.
3
Select the discriminant range.
- Press [1]-[2] keys to go to <133 RANGE>
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
5: +5 (e/d) or more
FULL: Entire area
- Press [4] key to fix.
4
Select the discriminant method.
- Press [1]-[2] keys to go to <134 METHOD>
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
ABSOL: Absolution value method
RELAT : Relative value method
- Press [4] key to fix.
5
Set HI buzzer
- Press [1]-[2] keys to go to <135 HI BUZZER>
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
OFF: OFF
ON : Active when above upper limit
- Press [4] key to fix.
6
Set OK buzzer
- Press [1]-[2] keys to go to <136 OK BUZZER>
- Press [4] key to change the setting value.
- Press [1]/[2] key to select.
OFF: OFF
ON : Active when discriminant result
is OK
- Press [4] key to fix.