64
5
The span adjustment starts.
For HJ17K0.1S(R) and HJ22K0.1S(R):
For HJ33K0.1S(R), and HJ62K0.1DS(R):
- Place the weight in the centre of the
weighing pan.
・
For HJ17K0.1S(R) and J22K0.1S(R):
Display changes to the order of
<on FS> → “blinking of <on FS>”.
Then span adjustment starts.
・
For HJ33K0.1S(R), HJ62K0.1DS(R):
Display changes to the order of
<on FS> → <push M>.
- Press [Menu] key.
Display changes to “blinking of
<on FS>”, then span adjustment
starts.
On completion of the span adjustment,
the display automatically changes to
<631 EX CAL>.
- Unload the weight from the weighing
pan.
6
Exit the setting menu.
- Press [Menu] key to shift to the
measuring mode.
(1) List of weights used for the span adjustment by model (Unit: g).
Model
HJ17K0.1S(R) HJ22K0.1S(R) HJ33K0.1S(R)
HJ62K0.1DS(R)
Selectable
weight on
the menu
17000
22000
33000
62000
set
1 - 17000
1 - 22000
1 - 33000
1 - 62000
(2) The span adjustment by the use of a weight less than the weighing capacity may possibly indicate
<UC> on the display. When this is the case, the weighing accuracy is not guaranteed.
Conditions under which <UC> is indicated;
-
-
When a sample that is more than two times heavier than
the weight that was used for the span adjustment is
weighed, and
When the readability (minimum indication) setting
<64A READABILIT>, which is finer than the minimum
interval for rounding setting <SELECT MIN> selected for
the span adjustment, is performed.