68
For HJ33K0.1S(R), and HJ62K0.1DS(R):
・
For HJ33K0.1S(R), and HJ62K0.1DS(R):
Display changes to the order of
<on FS> → <push M>.
- Press [Menu] key.
Display changes to “blinking of
<on FS>”, then span adjustment starts.
On completion of the span adjustment, the
display automatically changes to <on 0>.
- Unload the weight from the weighing pan.
Display automatically changes to “blinking of
<on 0>” and zero-point adjustment starts.
On completion of the zero-point adjustment, the
display automatically changes to “blinking of
<Ch 0>”.
4
The calibrating the internal weight starts.
Display changes to the order of
“blinking of <Ch 0>”
→
<Ch FS>
→
“blinking of <Ch FS>.
On completion of the calibrating the internal
weight, the display automatically changes to
<636 REF CAL>.
5
Exit the setting menu.
- Press [Menu] key to shift to the measuring
mode.