Removing the rubber cap
If the balance takes longer than usual to be stabilized, the filter of the balance may be clogged.
In this case, remove the black rubber cap shown in the following figure for a quick fix. Force the
cap open with a flathead screwdriver or a pair of pliers (The cap is somewhat difficult to remove to
protect against dust and water).
If the problem is cleared by removing the rubber cap, replace the filter with a new one. If the
problem is not cleared by removing the cap, it has been caused by another reason. Contact our
Technical Service Division or a store dealing in our balances for either case.
Caution:
The balance no longer conforms to IP65 after the rubber cap is removed. When the cap is removed,
protect the balance against dust or water drops.
Rubber cap
Enlarged view