Maximum Input Voltage
250V dc or rms ac on all ranges
DIODE TEST/CONTINUITY
Maximum Test Voltage
≤2V
Test Current
Approximate 1mA when measuring forward bias junction.
Continuous Resistance
≤150Ω
FREQUENCY
Range
Resolution
Accuracy
300Hz
0.01 Hz
0.05%+2
3KHz
0.1 Hz
30KHz
1 Hz
300KHz
10 Hz
3MHz
100 Hz
30MHz
1KHz
Frequency Range
10Hz to 30MHz
Applicable Function
Volts AC
Sensitivity of AC Voltage
ENVIRONMENTAL
Temperature Coefficient
<0.1 times the applicable accuracy specification per degree C for 5
℃
to
18
℃
and 28
℃
to 40
℃
.
Operating Temperature
5
℃
to 40
℃
Storage Temerature
-25
℃
to 50
℃
Relative Humidity
≤90% at 5
℃
to 28
℃
(non-condensing)
≤80% at 28
℃
to 40
℃
≤70% at 10
℃
to 40
℃
for the 300KΩ, 3MΩ and 30MΩ ranges.
GENERAL
Common Mode Voltage
1000V dc or peak ac maximum from any input to earth
Size
245mm deep, 220mm wide, 82mm high
Weight
about 2Kg
Power
about 5VA
ACCESSORIES
A couple of test leads
A power line cord
A user’s manual
SECTION 7
Points for Attention to Use of Operation Instruction
The present operation instruction is subject to change without notice.
The content of the operation instruction is regarded as correct. Whenever any user finds its mistakes, omission,
etc., he or she is requested to contact the manufacturer.
FREQUENCY
LEVEL
(
SINE WAVE
)
10Hz~300KHz
>
300mV
300KHz ~30MHz
>
1.5V
21