GT ULTRA ZERO AIR GENERATOR
USER MANUAL
www.vicidbs.com
Version 1.03.0000
6.5 Unit Installation
The GT Ultra Zero Air Generator can be installed two different ways. Either placed on a surface, like a table or
lab bench, or mounted to a wall.
A. On a table/lab bench:
For optimum ventilation allow at least 150 mm (6 inches) clearance on either side of the unit.
B) Mounted to a wall:
• Drill holes 200 mm (7.87 inches) apart into the wall and
insert the supplied mounting screws. Tighten only enough
so there is approximately 1/4” extruding from the surface.
• Hold the generator up to the extruding screws and insert the
unit over the screws through the keyholes. Let the unit slide
downwards as the screws slide up into the keyhole slots.
WARNING: Do not tighten screws fully. It must be possible to remove the generator
from a permanent structure without the use of tools.