GT ULTRA ZERO AIR GENERATOR
USER MANUAL
www.vicidbs.com
Version 1.03.0000
CONDITION
Warm-up /
Check Pressure
Ready
Check
System
Number on
the Board
YELLOW
GREEN
RED
SYSTEM GENERAL:
Power Fault:
OFF
OFF
OFF
Fault: Catalyst replacement
After 3 years
(1)
Blink 1
Hz
4
After 3 years plus 60 days
OFF
ON
4
STARTUP:
Quick Start Mode (0 to 3 minutes max.)
System Fault: Catalyst temperature is low at the 40
min checkpoint
5
Warm-up Mode (3 to 40 minutes max.)
System Fault: System is not warming up
OFF
OFF
ON
5
System Fault: Catalyst temperature is low at the 40
min. checkpoint
OFF
OFF
ON
5
OPERATION:
System Fault: Catalyst temperature exceeds limit
of 650°C
OFF
OFF
ON
6
System Fault: Inlet pressure low (< 4.5 bar)
ON
ON
ON
1
System Fault: Inlet pressure high (> 8 bar)
ON
ON
ON
2
System can be reset to initial condition:
• Turn unit off to reset system
•
Turn unit on and check the LED configuration
9. TROUBLESHOOTING