8.3. CONFIGURE PARAMETERS
SELECTION
DESCRIPTION
OPTIONS
DEFAULT
Set Clock
Adjust the internal clock
-
-
Low Press Error
*
Percentage of total pressure that has to be
exceeded to activate the low pressure alarm
0-100%
30%
Pressure Rise
Sets how fast the pressure should increase.
If the pressure increases at a slower rate, the
low pressure alarm will activate.
0.01 to 4.2 bar/min 14
0.2 to 60 psi/min
0.09 bar
14 psi
Pressure Drop
Delay
Sets a delay in seconds to ignore a pressure
drop (override low pressure alarm)
2 to 10 minutes
2 minutes
Auto Start
After powering the generator, it goes back to
the status before power shutdown.
Exam ple:
1. System running, producing hydrogen
and valve is open
2. Power shutdown
3. Power restarted
4. System starts running and then opens
the valves automatically after reaching
the internal pressure
YES / NO
NO
Beeper
Sets whether the audible signal is activated
in the event of an alarm
ON / OFF
ON
Cascading Mode
Sets the cascading mode. (see cascading for
more details)
None / Normal /
Ext. P control
None
N Gen H2 in
Cascading
Sets the number of generators in the
cascading chain
1 to 10
1
ID Number
Sets the ID number
1 to 10
1
Remote Start/Stop
Mode
Configures the remote Start/Stop function
Start/Stop, Start only,
Direct control
Start/Stop
Auto Refill
If set to ON, the pre-level water alarm is
used to trigger an external pump or valve to
refill the water tank
ON / OFF
OFF
Auto Refill Timeout
Within this time the internal water tank
must be filled. If a time out occurs the alarm
“check auto refill” appears.
5-600 sec.
30 sec.
Remote Relay
Mode
Configures the remote relay contacts
Normally Open (NO)
Normally Closed (NC)
NC
Pre Alarms in List
If set to YES, the pre alarms are also shown
in the alarm log
YES / NO
NO
Pressure Units
Sets the desired unit of measure for the
pressure
bar / psi / kPa
bar