VC837 DIGTAL MULTIMETER Operation Manual

1.

General Description

This is a 3 1/2 digital multimeter with high stablility and performance. It uses a LCD with
20mm high figure, which makes the reading clearer and the operation more convenient.
It can test DCV, ACV, DCA, ACA, resistance, capacitance, frequency, NCV,

duty cycle,

temperature, diode, and continuity. This meter also designed with functions including
unit symbol display, data hold, lighting, auto range, auto power off and warning
functions. To assure high accuracy and resolution, it adopts integrated circuit 8‐bit
microprocessor and a dual integral A/D conversion as LCD driver, giving high resolution
and high accuracy. It is an ideal tool for labs, factories and radio‐technology.

2. Safety Instructions

The instrument is designed in compliance with IEC1010 standard (safety standard issued

by International Electro technical Committee). Please read the following safety

instructions before operation.

2.1 Check the connection and insulation of test leads to avoid electric shock.

2.2 To avoid electric shock and damage to the meter, do not input voltage exceeding

rated value.

2.3 When measuring voltage higher than DC 60V or AC 40V, please be careful and avoid

electric shock.

2.4 Select correct function and range to avoid wrong operation.

2.5 Move the test leads away from test points when switching to other function.

2.6 Don’t input voltage in current terminal.

2.7 Don’t make any modification to the circuit. It may damage the meter or jeopardize

safety.

2.8 Safety symbols:

” High voltage, “

” GND, “

” Dual insulation, “

” Refer to manual, “

” Low

battery indication.

3. Features

3.1 General Characteristics

3.1.1 Display: LCD;

3.1.2 Max display: 6000 (3 5/6 digits, automatic polarity, and unit symbol display);

3.1.3 Measurement method: Analog to digital converter (in micro processor ADC+MCU);

3.1.4 Sampling rate: approx.3 times/sec.

3.1.5 Over‐range display: “OL” displayed.

3.1.6 Low battery indicator: “ ”

3.1.7 Working environment: (0

40)

, relative humidity: <80%;

3.1.8 Store condition: (‐10

50)

, relative humidity: <80%

3.1.9 Battery: 2 pieces 1.5V battery (“AAA” 7# battery);

3.1.10 Dimension: 140×72×37mm (length x width x height);

3.1.11 Weight: approx. 195g (including battery);

3.1.12 Accessories: test leads, TP01 thermocouple, user manual, holster, gift box, and

2*1.5V batteries.

3.2 Technical Features

3.2.1 Accuracy: ± (a% × r digits). To assure accuracy, the ambient temperature

should be (23±5)

, relative humidity <75%. One year accuracy is guaranteed since

production date.

3.2.2 DC Voltage

DCV

Range

Accuracy

Resolution

600mV

±(0.5%+4)

0.1mV

6V

1mV

60V

10mV

600V

±(1.0%+4)

100mV

Input impedance: at 200mv range >40MΩ, at other ranges is 10MΩ.

Overload protection: 600V DC/AC rms.

3.2.3. AC Voltage

ACV

Range

Accuracy

Resolution

Summary of Contents for VC837

Page 1: ...n and range to avoid wrong operation 2 5 Move the test leads away from test points when switching to other function 2 6 Don t input voltage in current terminal 2 7 Don t make any modification to the c...

Page 2: ...onds Overload protection 0 5A 250V fuse and 10A 250V fuse 3 2 5 AC Current ACA Range Accuracy Resolution 600uA 1 5 5 0 1 A 6000uA 1 A 60mA 10 A 600mA 100 A 6A 1mA 10A 2 0 10 10mA Maximum voltage drop...

Page 3: ...1832 750 1 0 5 750 1 5 15 Thermocouple K type WARNING do not input voltage at this range Operation 4 1 Panel Description LCD Display the data and unit symbol Number Feature Description 1 Relative REL...

Page 4: ...e is no measurement in 15 minutes the meter will automatically power off and enter sleep mode One minute before sleep mode the buzzer will beep for 5 times to remind user Press any button or turn the...

Page 5: ...s the default when you turn on the meter AUTO symbol displayed Press RANGE REL key can change to Manual range the available ranges are 600uA 6000uA 60mA 600mA 6A 10A 4 4 3 Connect test leads to the te...

Page 6: ...ance value NOTE 1 Don t input voltage or current under capacitance ranges 2 Press RANGE REL to clear the reading before testing to assure the accuracy 3 There is only the auto range mode under the cap...

Page 7: ...RATURE MEASUREMENT 4 11 1 Turn the rotary switch to range Press SELECT key to select or mode 4 11 2 Insert the cold end free end of thermocouple in V mA and COM terminal and put the working end temper...

Page 8: ...ter does not work properly please check the meter as following steps If the problems still cannot be solved please refer to repairing center or contact the local dealers Fatult Solution No reading on...