6uF
1nF
60uF
10nF
600uF
±(5.0%+8)
100nF
6mF
1uF
60mF
10uF
Overload protection: 250V DC/AC peak value.
3.2.8 Frequency
Range
Accuracy
Resolution
10Hz
±(0.5%+4)
0.01Hz
100Hz
0.1Hz
1000Hz
1Hz
10kHz
10Hz
100kHz
100Hz
1MHz
1kHz
10MHz
10kHz
Input sensitivity: 0.7V rms.
Overload protection: 250V DC/AC peak value.
3.2.9
Diode and Continuity Test
Measurement
Range
Test conditions
Diode forward voltage drop.
Forward DC current is
approx
0.8mA,
reverse
voltage
is
approx 2.2V.
When the resistance under test
is less than 50Ω, buzzer sounds
continuously.
Open circuit voltage:
2.2V
Overload protection: 250V DC/AC peak value.
WARNING: Do not input voltage at this range!
3.2.10 Temperature (
℃
/
℉
)
Range
Accuracy
Resolution
‐40
℃
~1000
℃
<
400
℃
±(1.0%+5)
≥400
℃
±(1.5%+15)
1
℃
0F~1832
℉
<
750
℉
±(1.0%+5)
≥750
℉
±(1.5%+15)
℉
Thermocouple: K type.
WARNING: do not input voltage at this range!
Operation
4.1 Panel Description
①
. LCD: Display the data and unit symbol
Number
Feature
Description
1
△
Relative (REL) mode is active
2
Data Hold is active.
3
-
Indicates negative readings.
4
DC
DC voltage or current measurement.
5
AC
AC voltage or current measurement.
6
AUTO
Auto range mode.
7
Null
8
Diode test mode.
9
The continuity beeper is on.
10
MAX MIN
Null
11
APO
Auto power off symbol.
12
The symbol of high voltage risk indication.
13
Low battery indication. Warning: To avoid error readings, which could lead to
possible electric shock or personal injury, please replace the battery in time.
14
h
FE
%,℃,℉
MΩ,kΩ,Ω
Hz,kHz,MHz
mV, V
uA, mA, A
Null;
Null;
Duty cycle, Degrees Celsius, Degrees Fahrenheit;
Megohm, Kilohm, Ohm;
Hertz, Kilohertz, Megahertz;
Millivolts, Volts;
Microamp, Milliamp, Amperes (A).
②
. Rotary switch: it is used to change the range and choose functions.