8 Technical Drawings
M12 Sensor
Connector
Mounting Holes
M4, depth 8
9 CE Compliance
Compliance with DIN EN 61326-1: 2006
EN 61000-4-3
EMC, Radiated radio-frequency - Electromagnetic field
immunity, Test Result A, Test with 10 V/m (0.15 ... 1000 MHz)
EN 61000-4-4
EMC, Electrical fast transient/burst immunity test, Test Result A
(see restrictions below)
EN 61000-4-6
EMC, Immunity to conducted disturbances, induced by
radiofrequency fields, Test Result A, Test with 10 V/m
EN 55011
Electromagnetic disturbance characteristics, Limit 30
dBμV/m
For testing typical settings for bubble detection have been applied.
The immunity depends on a reasonable configuration.
Operating with very high bubble sensitivity combined with a very short response time
can cause disturbances of the system, induced by
electromagnetic disturbing, pressure changing, mechanical vibration, etc.