MOTOR DRIVER
COMMUNICATION *
* Iris – Rotation – Shutters
The collimator motor does not
communicate with the central unit.
Press the “ALARM RESET” key in
order to go on to work (it is possible to
go on with XR, if the I.I. field results to
be free).
In case the error persists, call Service.
FILAMENT ERROR
Problems found in the filament
management.
Press the “ALARM RESET” key in
order to go on to work.
If the error repeats and persists also at
the next start-up, call Service.
KV ERROR
Error in the kV generation
Press the “ALARM RESET” key in
order to go on to work.
If the error repeats and persists also at
the next start-up, call Service.
mA OVERLOAD
Error in the mA reading (over the
max. allowed limit)
Press the “ALARM RESET” key in
order to go on to work.
If the error repeats and persists also at
the next start-up, call Service.
THERMIC ALARM
The temperature of the monobloc has
reached the max. allowed value.
Wait for the tube cooling
EMERGENCY UP/DOWN
PUSHED
Pressed UP/DOWN emergency push-
button.
Reset the correct working of the
emergency push-button by turning it
clockwise.
POWER SUPPLY ERROR
Incorrect secondary power supply
Switch the unit OFF, wait for some
seconds and switch ON again.
In case the error persists, call Service.
FOOTSWITCH OR
HANDSWITCH ERROR
Faulty or damaged footswitch or
handswitch for the x-ray control
Disconnect and reconnect the
footswitch.
In case the error persists, call Service.
STARTER FAULT
Only rotating anode
Error in the circuit of the rotating
anode.
Switch the unit OFF, wait for some
seconds and switch ON again.
In case the error persists, call Service.
OPEN DOOR WARNING
For the units provided with the door
control, the door to enter the room is
open.
Check that the door is closed
correctly.
In case the error persists, call Service.
VIDEO SIGNAL NOT OK
The video signal is not present or it is
not bright enough.
Switch the unit OFF, wait for some
seconds and switch ON again.
Check that the trolley is connected to
the unit and that the BNC cables are
properly connected to the monitors.
In case the error persists, call Service.
UNAVAILABLE INVERTER
POWER SUPPLY
Unavailable inverter power supply
Switch the unit OFF, wait for some
seconds and switch ON again.
In case the error persists, call Service.
MAX. X-RAY TIME
The max. exposure time has been
reached.
Press “ALARM RESET” key to go on,
repeat x-rays.
In case the error persists, call Service.
TIMEOUT
The preparation handswitch is held
down for more than 15 seconds
without performing x-rays.
Release the preparation handswitch
and repeat the operation.
MANUAL X-RAY STOP
The x-ray hand switch has been
released before the end of the
exposure.
Press “ALARM RESET” key to go on
and repeat the exposure.
MISSING PULSES
Control pulses lack from the memory
Press the “ALARM RESET” key in
order to go on to work.
In case the error persists, call Service.
CCA BUSY
Busy central unit control
Press the “ALARM RESET” key in
order to go on to work.
In case the error persists, call Service.