3.6.3. AUTOMATIC AND MANUAL PULSED FLUOROSCOPY
Selection of the PULSED FLUOROSCOPY frequency. By pressing the key, the
selection and the indication aside change (according to the installed memory).
Press the right pedal to control the exposure.
After the kV balance time, the exposure with the frequency set on the display is
performed (the frequency changes according to the features of the installed
memory).
In MANUAL mode, the kV and mAs value must be set by the operator. The
exposure parameters variation allows to modify the image quality at the discretion
of the operator.
X-ray emission LEDs on the keyboard and on the monitor trolley lit.
3.6.4. “SNAP SHOT”
No selection from the keyboard must be performed.
The “Snapshot” can be performed in Automatic and Manual Fluoroscopy mode
without performing any additional choice on the keyboard.
Press the pedal of the X-ray control footswitch.
After the kV balance time, an exposure of about 1 sec controlled directly from the
unit is performed. This exposure allows to get an image that is at high contrast
and without background noise.
X-ray emission LEDs on the keyboard and on the monitor trolley lit.