User's Manual
ARCOVIS 3000 S/R
Villa Sistemi Medicali
Pag. 34/76 - Rev. F
[File:201175-G-01-01.doc]
3.6.5. RADIOGRAPHY
Select
Radiography
. The display appears as in the figure below:
Key for the FLUOROSCOPY mode selection (§3.6.1)
Key for the APR 1 mode
selection. (Head)
Key for the APR 2 mode
selection. (Spinal column)
Key for the APR 3 mode
selection. (Pelvis)
The chosen APR program proposes the default
exposure parameters (kV, mAs, focus). The
proposed data can be modified according to the
requirements.
HiPower Radiography selection
key
It appears only if the set mAs are lower than
2,2mAs.
Put the cassette holder at the input of the I.I. tube by coupling the two fixed locks
and by locking the knob, insert the loaded x-ray cassette.
Set the kV and mAs parameters by operating on the relative increase and decrease
keys placed under the indication of the value selected on the display.
Use the whole length of the extensible cable to move away as much as
possible from the irradiated area
.
Press the x-ray handswitch and hold it down till the end of the exposure.
X-ray emission LED on the keyboard lit.
At the end of the exposure the performed exposure time is displayed.