APR radiography
By selecting the wished APR program, the relative key appears in negative and, on
the right side of the screen, the keys concerning the selection of “slim patient” and
“sturdy patient” appear.
Indication of the selected APR program.
Key to select the APR values for slim patient (visible only if an APR program is
selected).
Key to select the APR values for sturdy patient (visible only if an APR program is
selected).
When the prearrangement for slim or sturdy patient is selected, the icon of the key
in “negative” appears.
Data storage key
The proposed data can be modified at pleasure
by the user. If the storage function is enabled in
the USER SETUP menu (§3.6.10), the modified
values can be saved in memory through the
relative key. In case the modified data are not
stored, at next return in Radiography mode, the
default data will be re-proposed.
HiPower radiography
By selecting HiPower radiography, the relative key appears in negative, the keys
concerning the APR programs disappear and the mAs are limited to a value lower
than 2,2mAs. The kV can be set without limitations.
In this mode it is possible to perform a radiography at 5kW with rotating anode and
at 3,5kW with fixed anode.